| Type | Face velocity (lfpm) |
Airflow Pattern | Radionuclides/ Toxic Chemicals |
Biosafety Level(s) | Product Protection |
| Class I * open front | 75 | In at front; rear and top through HEPA filter | No | 2,3 | No |
| Class II Type A | 75 | 70% recirculated through HEPA; exhaust through HEPA | No | 2,3 | Yes |
| Type B1 | 100 | 30% recirculated through HEPA; exhaust via HEPA and hard ducted | Yes (Low levels/volatility) |
2,3 | Yes |
| Type B2 | 100 | No recirculation; total exhaust via HEPA and hard ducted | Yes | 2,3 | Yes |
| Type B3 | 100 | Same as IIA, but plena under negative pressure
to room and exhaust air is ducted |
Yes | 2,3 | Yes |
| Class III | NA | Supply air inlets and exhaust through 2 HEPA filters | Yes | 3,4 | Yes |
* Glove panels may be added and will increase face velocity to 150 lfpm; gloves may be added with an inlet air pressure release that will allow work with chemicals/radionuclides.